Precautions when using Kyushu space biochemical incubator

九州空间生化培养箱使用时的注意事项

Installation and commissioning of biochemical incubator

1. Be careful not to use force on the glass during transportation, and do not hit the cabinet. Inverting and tilting greater than 45° are prohibited.

2. After the equipment is grounded, it should be placed as stable as the ground is uneven.

3. Please do not use the instrument in direct sunlight or high temperature and humidity. The ambient temperature should be kept at 10-30 °C.

4. This equipment should be kept away from electromagnetic interference sources, and the grounding wire of the equipment should be effectively grounded.

5. When the equipment is in normal operation, the load placed in the box should not affect the air circulation to ensure the air circulation and temperature uniformity inside the box.

6, the power supply is AC 220V50HZ, must use 10A three-core socket, and has a reliable grounding wire, you must not use the two-core socket.

7, the power cord should not be close to the back, do not let the instrument or other items on the power line, so as not to damage the power cord

Precautions when using biochemical incubator

1. Do not put flammable and volatile chemicals into the box.

2. If abnormality, odor, smoke, etc. occur during use, please turn off the power immediately. Users should not blindly repair it. They should inform the repair department of the company and check it out by professionals.

3. The wall of the tank wall and the surface of the equipment should be wiped frequently to keep it clean and increase the transparency of the glass. Do not use acid, alkali or other corrosive solutions to wipe the outer surface.

4. When the equipment is not in use, it should be treated with moisture. The specific method is as follows:

Pour the water from the bottom of the tank to the water tray, set the temperature to 42 ° C, run for 5 hours, and open the door every two hours to release the moisture. After the treatment is finished, unplug the power plug and store it.

5. If the device is not used for a long time, unplug the power cord to prevent the device from damaging the person. It should be operated regularly (usually in the first quarter) for 2-3 days according to the conditions of use to remove moisture from the electrical parts and avoid damage to the relevant components.

Twist Car

twist car

Baby twist car is non electronic ride on toy, by swiveling right and left to move. More healthy than the electronic ride-on, it is sport baby toy for kids who are more than 3 years old.

Twist car series have 4 different shapes, two classic plasma cars and two new ones. Two classic ones are keeping production more than ten years, are our hotsellers. Two new ones are designed by Italian designers, unique all over the world and suitable for high level.

Howawa is top twist car manufacturer in China, provide OEM & ODM service, all the twist cars are made of new PP material, get CE and ISO9001, pass European EN71 & American ASTM tests. Have three wheel choices: PP wheels, PU wheels, and PU light wheels. And two packing way choice: 1pc in 1 box and 4pcs in 1 box.

Twist Car,Kid Twist Car,Wiggle Car,Swing Car

​Taizhou Howawa Baby Products Co., Ltd. , https://www.howawa.com

Posted on